Clinical Outcomes of the Sentio Active Transcutaneous System
Clinical Outcomes of the Sentio Active Transcutaneous System
A High-Performance Direct-Drive Solution
Following the recent publication of our multicenter clinical investigation, we’re pleased to share a deeper look at the evidence supporting the Sentio™ System’s safety, audiological performance, and positive patient-reported outcomes. This study provides compelling confirmation of the Sentio System’s ability to deliver consistent safety, meaningful audiological benefits, and high levels of patient satisfaction.
Key Takeaways:
- The 3- and 6-month audiological performance, safety, and subjective benefit outcomes of the Sentio System have been published in Otology and Neurotology.
- Significant improvements in soundfield thresholds, as measured by functional gain, averaged 32.8 dB and significant improvements in speech-in-quiet and speech-in-noise were reported.
- Mean surgery time was 58 minutes and 100% of implant sites had healed within two weeks.
- The majority of patients reported excellent satisfaction and demonstrated substantial quality of life (QoL) improvements. These same patients also reported an average wear time of greater than 8 hours of daily device use.
- Safety was confirmed as no serious device-related adverse events occurred. Among a limited number of adverse events—all of which were mild to moderate in nature.
Study Design and Population
This prospective multicenter investigation enrolled 51 adult patients (≥18 years; all unilaterally implanted) across six European tertiary centers. Indications included CHL (51%), MHL (25%), and SSD (24%). Each patient served as their own control with assessments done prior to implantation (baseline) and after implantation at 2, 4, and 6 weeks, and 3, 6, 12, and 24 months. Primary outcomes were assessed at 3 months after implantation with confirmatory analyses done at 6 months (PPS, n=32) after implantation.
Surgical and Postoperative Findings
Several surgery-related outcomes were measured, for example, the type of anesthesia used and the duration of surgery. Postoperative wound healing as well as patient-reported pain and numbness were also assessed over follow-up visits. Key findings:
- Mean surgical time: 58 minutes, trending shorter over time
- Anesthesia: 96% under general anesthesia; 4% under local anesthesia.
- Wound healing: 100% healed within two weeks
- Pain and numbness: Minimal pain and numbness reported
- Preoperative imaging: 55% of implantations were done without preoperative imaging
- Intraoperative events: No cases of dura exposure, excessive bleeding, or aborted surgeries due to anatomic challenges
These findings underscore the Sentio System’s anatomical versatility and streamlined procedural workflow.
Audiological Outcomes
The Sentio System delivered powerful and meaningful improvements across CHL, MHL, and SSD indicated patients:
The clinical results demonstrate that the Sentio System delivers powerful and consistent audiological improvements across conductive, mixed, and single-sided deafness indications. Patients achieved an average functional gain, a measure of improvement in sound-field thresholds from unaided to aided conditions, of 32.8 dB. Effective gain, on the other hand, compares aided sound-field thresholds to the patient’s own bone-conduction thresholds. In CHL/MHL users, this value—seen as the remaining air-bone gap—is an indicator of how completely the direct-drive system is overcoming conductive components of hearing loss. With Sentio, the effective gain demonstrated a residual gap of just 8.96 dB, underscoring precise mechanical coupling and highly efficient vibration delivery.
Speech outcomes were equally impressive: aided speech recognition in quiet reached 97.9%, and speech-in-noise testing showed a −2.91 dB improvement in SNR compared to preoperative softband performance, with an even greater benefit (−5.41 dB) for SSD patients. Together, these results highlight the Sentio System’s ability to provide clear, natural sound and measurable improvements that translate directly to real-world communication.
Functional Gain
- Functional gain is a measure of the difference between unaided and aided sound field thresholds.
- Average functional gain (PTA4) measured at 3 months: 32.8 dB (p < 0.001)
- Stable functional gain was maintained at 6 months
Effective Gain
Effective gain is the difference between aided sound field thresholds and unmasked
bone conduction thresholds (applicable for CHL/MHL indications only)
- Average remaining ABG of 8.96 dB (p < 0.001) measured at 3 months.
Speech Recognition
- Speech recognition in quiet improved by 51.6 percentage points, reflected in a mean aided score of 97.9%.
- Speech-in-noise performance improved by −2.91 dB SNR compared with preoperative test performance
- SSD patients demonstrated an even greater benefit of the Sentio System in noisy listening conditions, with an average of −5.41 dB SNR improvement
These improvements translate into clinically significant real-world benefits, enabling clearer, more natural communication in both quiet and noisy environments.
Patient-Reported Outcomes (PROs)
Patients reported meaningful improvements in quality of life and high daily usability, reinforcing the Sentio System’s strong real-world acceptance. Key findings include:
- Glasglow Benefit Inventory (GBI): At 3 months, the mean score was +29.2, indicating a clinically meaningful improvement in overall quality of life. Ninety-six percent of patients reported a quality-of-life improvement
- Speech, Spatial, and Qualities of Hearing Scale (SSQ) 12: At 3 months, 100% of patients reported an improvement in their quality of hearing, with an average improvement of 2.8 points on a 10-pt scale
- International Outcome Inventory Hearing Aids (IOI-HA): At 6 months, most patients (92%) reported being highly satisfied with the aided performance with Sentio.
- Sound quality: 80% reported the sound quality as “natural”
- Retention and comfort: High patient ratings on retention and comfort
These patient-reported outcomes complement the audiological findings, demonstrating that the Sentio System delivers not only measurable hearing improvements but also enhanced quality of life and user satisfaction.
Safety Profile
The Sentio System demonstrated an excellent safety profile through the 6 months of published data in the study. A total of 56 adverse events (AEs) occurred among 25 patients and all were classified as mild to moderate and were consistent with the expected postoperative recovery process. Most AEs involved short-lived symptoms such as neck pain, headache, dizziness, or minor wound-related irritation, while soft-tissue issues were minimal and resolved without sequelae.
Key Safety highlights:
- No serious device-related adverse events
- 56 AEs among 25 patients, all rated as mild or moderate
- Most events were not unexpected in the initial postoperative period
Clinical Implications
The Sentio System’s small implant footprint and flexible surgical approach enhance clinical usability, while its robust audiological outcomes position Sentio as a strong choice for CHL, MHL, and SSD candidates.
The Sentio System delivers:
- Excellent hearing outcomes (functional and effective gain, speech recognition in quiet and in noise)
- Straightforward surgical workflow and short operating times
- Low complication rates and high patient satisfaction
Conclusion
This multicenter prospective investigation showed that the Sentio System is a safe, effective, and patient-centered solution for bone conduction hearing rehabilitation. With significant audiological gains, improvements in patient quality of life, and a strong safety profile, the Sentio System represents an advance in modern bone conduction technology.
References
-
Hol, Myrthe Karianne Sophiea,*; Aukema, Tjerk Winusa; Ubbink, Sander Wilhelmus Johannesa; Lindholm, Dennisb; Holmberg, Marcusb; Arndt, Susanc; Wolf, Stefaniec; Monksfield, Peterd; Marwa, Mihalached; Lenarz, Thomase; Busch, Susane; Tysome, Jamesf; Asher, Abif; Kok, Hermang; Mylanus Emmanuel Antonius, Mariag. A Prospective, Multicenter Clinical Investigation on the Safety and Performance of a New Active Transcutaneous Bone-Anchored Implant System. Otology & Neurotology ():10.1097/MAO.0000000000004706, November 17, 2025. | DOI: 10.1097/MAO.0000000000004706
-