Safe 1.5T MRI Scanning with the Sentio Ti Implant

Safe 1.5T MRI Scanning with the Sentio Ti Implant

Hospital radiology room with 3d rendering mri scanner machine

Key Takeaways

  • The Sentio™ Ti implant is 1.5T MR Conditional—no magnet removal or bandaging needed
  • Internal design reduces discomfort and ensures scanning safety for up to 30 minutes (as per the FDA)
  • Like all active transcutaneous implants, implant produces a localized artefact during MRI
  • If high-resolution imaging near the implant is likely to be required, a percutaneous bone anchored hearing system (BAHS) may offer clinical advantages
  • 3T MRI is not currently approved for Sentio; most MRI centers in the U.S. primarily use 1.5T systems

When recommending implantable BAHS, understanding MRI compatibility is an important part of long-term patient counseling. This includes both the safety of MRI scans and how the presence of an implant may affect image quality near the implant site. The Sentio Ti implant has been engineered with these considerations in mind.

Safe, Comfortable MRI at 1.5 Tesla

The Sentio Ti implant is MR Conditional at 1.5T and does not require magnet removal, bandaging, or a splint before scanning. This is made possible by a combination of engineering features, including a:

  • Titanium-encased retention magnet laser-welded to the transducer housing via a titanium reinforcement wire
  • Hardened silicone over-mold that distributes pressure evenly and resists migration
  • Subperiosteal antenna placement that adds mechanical stability and prevents dislodgement

These design elements were extensively tested in bench models, simulations, and clinical studies. The retention magnet’s movement is reduced by a factor of five compared to traditional designs, and the pressure pain safety margin is increased sixteenfold.1

In a pre-approval study, 15 patients underwent voluntary head MRI at 1.5T. All scans were completed successfully. On a scale of 0–10, where 0 indicated no sensation and 10 indicated the worst imaginable pressure or pain, the median pressure score was 3 and the median pain score was 1.2 No complications or scan terminations occurred, and radiographers noted no challenges when performing the scan.2

Understanding Image Artefacts

While 1.5T scans can be safely performed with the implant in place, it is essential to differentiate MRI compatibility from diagnostic utility.

MRI artefacts are caused primarily by the presence of the internal retention magnet. These signal voids (or shadows) can interfere with imaging around the implant site. The Sentio Ti implant produces an artefact radius of 18 cm. All active transcutaneous devices create a shadow that should preclude patients with neurological issues, head trauma, or other medical conditions. For patients who might require frequent or detailed head imaging, clinicians may consider recommending a percutaneous system like Ponto™ to ensure greater visibility of surrounding structures. The Ponto percutaneous implant has the smallest MRI artefact of bone anchored implants at just 1.5 cm.

Why Not 3T?

The Sentio Ti implant is not currently approved for use in 3T MRI scanners. While 3T MRI offers greater spatial resolution and signal-to-noise ratio, it also increases artefact size and poses higher safety demands for implanted magnetic devices.

However, 1.5T remains the standard of care for most brain and head/neck imaging protocols in the United States.3,4

  • As of 2023, approximately 60 percent of MRI scanners in the U.S. are 1.5T3
  • 1.5T systems offer sufficient diagnostic clarity for the majority of clinical cases, particularly outside of research settings
  • Many institutions rely on 1.5T as their default strength for routine neuroimaging

Designed for Everyday Life: Why We Chose a Stronger Magnet

When designing the Sentio Ti implant, we made a deliberate choice to prioritize exceptional retention in everyday life. This led to the inclusion of a stronger internal magnet, which helps ensure that the external sound processor stays securely and comfortably in place during daily activities. While we recognize that stronger magnets can result in larger MRI artefacts, we believe that for most patients, the benefits of enhanced retention, stability, and comfort far outweigh the impact on these diagnostic scenarios, especially considering that scans of the implant area can often be addressed with alternative imaging strategies. This design decision reflects our commitment to optimizing long-term wearability and user experience.

For full MRI guidelines, technical specs, or patient counseling tools, reach out to your Oticon Medical representative. It’s important to note that the Sentio 1 Mini sound processor should be removed prior to MRI scanning.

 

References:

  1. Hanna R, et al. “MRI Safety of the Sentio Ti Implant.” Internal Study Data, 2024.
  2. The Sentio System – Almost nothing to see, so much to experience, Whitepaper, 2024, Doc 282508
  3. IMV Medical Information Division. MRI Market Summary Report. 2023.
  4. Kanal E, et al. ACR Guidance Document on MR Safe Practices. J Magn Reson Imaging. 2021;53(6):1718–1727.
  5. Cochlear Osia System MRI Guidelines; MED-EL Bonebridge Clinical Information.